HT-9464 High Voltage Tester

products > high voltage testing > high voltage testing comparison

HT 9460 and 9464 High Voltage Isolation Test Systems Comparison

  HT 9464 Specifications HT 9460 Specifications
Voltage Range 0 to 7.00 kVac or 0 to 9.99 Kv peak with 10 Volts resolution. Accuracy of 2% +/- one digit. 0 to 4.0 kVac or 5.66 volts peak for handler configuration of partial discharge testing at 5 pC. 0 to 7.00 kVac, or 9.99 peak volts, 10 Vac resolution. Accuracy of 2% +/- one digit.
Test Voltage Duration T1 and T2 from 0.2 to 99.9 seconds. 200 seconds max total. 0.2 to 99.9 seconds.
Programmable Leakage Test Threshold 0 to 99 micro-amps in 0.1 micro-amp steps, 100 meg ohm in parallel with 10 pF current limit. Accuracy of 2% +/- one digit. 0 to 50 micro-amps in 0.1 micro-amp steps, 150 Meg ohm current limiting resistor. Accuracy of 2% +/- one digit.
Partial Discharge Test Detection Single event or digitally multiple events.  
Partial Discharge Test Threshold and Delay< 0 to 50 picocoulombs in 0.1 pC steps with 0.0 to 9.9 second delay.  
Display Mode and 3-1/2 Digit Panel Meter Source or Device Voltage (kVac or kV peak), Leakage Current (micro-amps) Source or device Voltage (kVac), Leakage Current (micro-amps) or blank display.
High Voltage 0 to 7.0 kVac or 0 to 9.99 kV peak, accuracy 2% +/- one digit. 0 to 7.0 kVac or 0 to 9.99 kV peak, accuracy 2% +/- one digit.
Low Limit Threshold Ensures that the actual device voltage is not below requirement. Ensures that the actual device voltage is not below requirement.
LED Indicators Testing, Pass, Fail High Voltage, Fail Leakage and Fail Corona. Testing, Pass, Fail High Voltage, Fail Leakage and Fail Corona.
Audible Alarm. 100 ms alarm for each device fail, disable through panel switch. 100 ms alarm for each device fail, disable through panel switch.
Test Counter 6 Digit Counter for total number of devices passed. 6 Digit Counter for total number of devices passed.
Failure Counter 6 Digit Counter for total number of device failures. 6 Digit Counter for total number of device failures.
Output Signals Handler Interface, Half Signal Marker, Test-in-Progress, Leakage monitor (1 micro-amp = 100 mv), Half Cycle with Corona. Handler Interface, Half Signal Marker, Test-in-Progress, Leakage Monitor (1 micro-amp=100 mv), Half Cycle with Corona.
Safety Interlocks Test Voltage Circuitry is automatically interrupted on opening handler device test area or removal of high voltage assembly. Test Voltage Circuitry is automatically interrupted on opening handler device test area or removal of high voltage assembly.
Operating Environment 15 to 35 deg. C; 35% to 75% relative humidity, non-condensing. 15 to 35 deg. C; 35% to 75% relative humidity, non-condensing.
Dimensions and Weight Manual version is 38.1 cm (15 inches) high, 30.5 cm (12 inches) wide, 81.2 cm (32 inches) long and weighs 31.8 kg (70 lbs). Auto version weight 31.8 kg (70lbs), size consult factory. Manual version is 38.1 cm (15 inches) high, 30.5 cm (12 inches) wide, 81.2 cm (32 inches) long and weighs 31.8 kg (70 lbs). Auto version weight 31.8 kg (70lbs), size consult factory.