HT 9464 Specifications |
HT 9460 Specifications |
Voltage
Range |
0
to 7.00 kVac or 0 to 9.99 Kv peak with 10 Volts resolution.
Accuracy of 2% +/- one digit. 0 to 4.0 kVac or 5.66 volts peak
for handler configuration of partial discharge testing at 5
pC. |
0
to 7.00 kVac, or 9.99 peak volts, 10 Vac resolution. Accuracy
of 2% +/- one digit. |
Test
Voltage Duration |
T1
and T2 from 0.2 to 99.9 seconds. 200 seconds max total. |
0.2
to 99.9 seconds. |
Programmable
Leakage Test Threshold |
0
to 99 micro-amps in 0.1 micro-amp steps, 100 meg ohm in parallel
with 10 pF current limit. Accuracy of 2% +/- one digit. |
0
to 50 micro-amps in 0.1 micro-amp steps, 150 Meg ohm current
limiting resistor. Accuracy of 2% +/- one digit. |
Partial
Discharge Test Detection |
Single
event or digitally multiple events. |
|
Partial
Discharge Test Threshold and Delay< |
0
to 50 picocoulombs in 0.1 pC steps with 0.0 to 9.9 second delay. |
|
Display
Mode and 3-1/2 Digit Panel Meter |
Source
or Device Voltage (kVac or kV peak), Leakage Current (micro-amps) |
Source
or device Voltage (kVac), Leakage Current (micro-amps) or blank
display. |
High
Voltage |
0
to 7.0 kVac or 0 to 9.99 kV peak, accuracy 2% +/- one digit. |
0
to 7.0 kVac or 0 to 9.99 kV peak, accuracy 2% +/- one digit. |
Low
Limit Threshold |
Ensures
that the actual device voltage is not below requirement. |
Ensures
that the actual device voltage is not below requirement. |
LED
Indicators |
Testing,
Pass, Fail High Voltage, Fail Leakage and Fail Corona. |
Testing,
Pass, Fail High Voltage, Fail Leakage and Fail Corona. |
Audible
Alarm. |
100
ms alarm for each device fail, disable through panel switch. |
100
ms alarm for each device fail, disable through panel switch. |
Test
Counter |
6
Digit Counter for total number of devices passed. |
6
Digit Counter for total number of devices passed. |
Failure
Counter |
6
Digit Counter for total number of device failures. |
6
Digit Counter for total number of device failures. |
Output
Signals |
Handler
Interface, Half Signal Marker, Test-in-Progress, Leakage monitor
(1 micro-amp = 100 mv), Half Cycle with Corona. |
Handler
Interface, Half Signal Marker, Test-in-Progress, Leakage Monitor
(1 micro-amp=100 mv), Half Cycle with Corona. |
Safety
Interlocks |
Test
Voltage Circuitry is automatically interrupted on opening handler
device test area or removal of high voltage assembly. |
Test
Voltage Circuitry is automatically interrupted on opening handler
device test area or removal of high voltage assembly. |
Operating
Environment |
15
to 35 deg. C; 35% to 75% relative humidity, non-condensing. |
15
to 35 deg. C; 35% to 75% relative humidity, non-condensing. |
Dimensions
and Weight |
Manual
version is 38.1 cm (15 inches) high, 30.5 cm (12 inches) wide,
81.2 cm (32 inches) long and weighs 31.8 kg (70 lbs). Auto version
weight 31.8 kg (70lbs), size consult factory. |
Manual
version is 38.1 cm (15 inches) high, 30.5 cm (12 inches) wide,
81.2 cm (32 inches) long and weighs 31.8 kg (70 lbs). Auto version
weight 31.8 kg (70lbs), size consult factory. |